Description
Overview
Every design decision in the B2 follows from one requirement: no air returns from the work zone to the room. The exhaust filter, the plenum, the bag-out port and the duct are all part of the containment boundary, and the cabinet is certified as a system rather than as a box.
That makes the installation a project. You need a dedicated duct sized to the exhaust volume, make-up air to match it, and a room that stays negative to the corridor at full flow. A B2 cabinet on an undersized duct will alarm continuously and will not hold containment, however good the cabinet is.
WSHLab builds the B2 from 4 ft to 6 ft, with a pressure sensor that holds total exhaust as the filters load, and a bag-out arrangement for the exhaust filter so service does not open the contaminated plenum to the room.
Key features
- 100% total exhaust. No recirculation path exists. Everything that enters the work zone leaves through the exhaust ULPA filter and the duct.
- Bag-out filter access. The exhaust filter is changed through a bag-out port on the contaminated side, protecting the service engineer and the room.
- Pressure sensor and duct static alarm. The cabinet monitors plenum pressure and duct static separately, so a blocked duct alarms before containment is lost.
- Cytotoxic-rated configuration. UL 1805 and UL 586 configurations, stainless work zone and a coved sump for spill control.
Specifications
| Airflow class | Class II Type B2 — 100% exhaust, no recirculation |
|---|---|
| Downflow velocity | 0.33 ± 0.025 m/s |
| Inflow velocity | 0.53 ± 0.025 m/s |
| Exhaust volume | 1200 m³/h (4 ft) to 1900 m³/h (6 ft) |
| Duct connection | Hard duct, sized to the exhaust volume |
| Filter | ULPA U15, 99.9995% at 0.12 µm, supply and exhaust |
| Working widths | 4 ft / 1.2 m, 6 ft / 1.8 m |
| External dimensions | 1100-1800 × 775 × 2250 mm (W × D × H) |
| Internal dimensions | 940-1625 × 600 × 660 mm (W × D × H) |
| Power | 1800 VA maximum, 230V or 115V single phase |
| Noise | ≤65 dB(A) |
| Certification | NSF/ANSI 49, UL 1805, UL 586, CE, UKCA, EN 12469 |
Where you would install it
- USP 800 compounding. Hazardous drug preparation with total external exhaust.
- Cytotoxic API handling. Powder weighing and dispensing where particles must not return to the room.
- Volatile toxic chemicals. Solvent and reagent work that cannot be recirculated.
Frequently asked questions
Can a canopy connection make an A2 cabinet equivalent to B2?
A canopy captures the exhaust but the cabinet still recirculates internally, so it is not equivalent. For USP 800, specify B2.
What duct size do I need?
Size for the exhaust volume on the datasheet — 1200 m³/h for the 4 ft model, 1900 m³/h for the 6 ft — and keep duct velocity within the range your HVAC engineer specifies.