Description

LICHEN LC-DBC-700IIA2 is a 700 mm desktop cabinet for space-constrained, low-risk work.

Marketplace listings and OEM catalogues are used for model and parameter research only. Final configuration, electrical supply, documentation, availability and pricing are confirmed by quotation.

Where this cabinet fits

Best for: Small labs, teaching benches and point-of-care rooms with limited space.

Not the right choice for: Two-person work, large equipment or chemical exhaust.

Specifications

Parameter Value
Brand / manufacturer LICHEN
Airflow 30% exhaust / 70% recirculation
External size 700×662×1260 mm
Workspace 498×570×530 mm
HEPA 99.995% @ 0.3 µm
Noise ≤67 dB
Power 1200 W
Weight 95 kg
Airflow class Class II Type A2, benchtop
Available sizes 700 mm
Compliance NMPA where applicable

Available configurations

Configuration Model Verified specifications
LC-DBC LC-DBC Size / configuration: 700 mm; Airflow class: Class II Type A2, benchtop

What to check before ordering

  • Confirm the airflow class against your work: Type A2 recirculates about 70% of the air, Type B2 exhausts 100% outdoors.
  • Measure the doorway, lift and bench run – a 6 ft cabinet needs clear access and a reinforced surface.
  • Confirm the electrical supply and whether a stabilised or dedicated circuit is available.
  • Decide who performs annual certification and how quickly HEPA filters can be replaced locally.

Additional information

Configuration

LC-DBC

Reviews

There are no reviews yet.

Be the first to review “LICHEN LC-DBC-700IIA2”

Your email address will not be published. Required fields are marked *

Technical documents

Datasheets, certificates and safety documents are available from our product specialists.

Request documentation