WSHLAB CO₂ Incubator

SKU LI-BPN-80-220V
Unit price · 80L · 220V/50Hz · 304 Stainless · 7-inch Touchscreen · RT+5~60°C · Yes · No
Price on request
Final price confirmed by quotationNo online payment required
Capacity: 80L
Voltage: 220V/50Hz
Chamber Material: 304 Stainless
Control Type: 7-inch Touchscreen
Temperature Range: RT+5~60°C
Humidity Control: Yes
Lighting Type: No
Total Price on request

Need bulk quantities or contract supply?Request a quotation with your required specifications.
Secure cart and checkout Verified listed-price options B2B quotation support Dispatch confirmed with your order

Compare every available specification below, or use the option selector above for a quicker purchase.

Cat Code CapacityVoltageChamber MaterialControl TypeTemperature RangeHumidity ControlLighting Type Price Qty Cart action
LI-BPN-80-220V 80L 220V/50Hz 304 Stainless 7-inch Touchscreen RT+5~60°C Yes No Price on request — Quote
LI-BPN-160-220V 160L 220V/50Hz 304 Stainless 7-inch Touchscreen RT+5~60°C Yes No Price on request — Quote
Total¥0.00

Showing 1 to 2 of 2 entries

Description

Overview

The BPN is the CO₂ incubator you specify when the cultures cannot be redone: a direct heat carbon dioxide incubator that holds 37°C and 5% CO₂ with an imported infrared sensor, and shows you the proof on a 7-inch touchscreen. Direct heat means the chamber walls themselves are the heating elements — recovery after a door opening is measured in minutes, with no water jacket to fill, treat, or eventually leak.

Everything inside the chamber is arranged for cell culture hygiene. The 304 stainless interior is pressed with rounded corners and a coved floor, so there is no crevice for a spill to hide in; the shelves and water dish lift out for autoclaving; and optional HEPA filtration and a UV-C decontamination lamp cover the multi-cell-line lab’s worst case. An SMC solenoid valve meters CO₂ from your cylinder, and the IR sensor reads concentration optically — immune to the humidity and temperature drifts that move thermal-conductivity sensors.

The controller logs temperature and CO₂ at one-minute intervals, replays the last hours as live curves, and exports CSV or PDF over USB — the medium pH story writes itself into your notebook. Two chambers sizes, 80L and 160L, cover a single-user plate workload to a multi-user culture room.

Key features

  • Direct heat design. heating elements bonded to the chamber walls recover RT+5~60°C quickly after door openings, with no water jacket maintenance
  • Imported IR CO₂ sensor. optical measurement of 0-20% CO₂, unaffected by chamber humidity — readings stay honest over long runs
  • SMC solenoid valve. precise gas metering from the cylinder to the chamber, with long service life
  • 304 stainless rounded-corner chamber. pressed interior with coved floor, wipes clean between cell lines; shelves and water dish autoclave-compatible
  • 7-inch touchscreen. real-time temperature and CO₂ curves, 1-minute-interval logging, CSV/PDF export over USB
  • Optional HEPA filtration and UV-C lamp. continuous air scrubbing and programmable decontamination for multi-line culture rooms
  • Over-temperature protection. independent alarm circuit guards unattended overnight and weekend runs

Specifications

Model range BPN-80 / BPN-160
Capacity options 80L · 160L
Temperature range RT+5~60°C, resolution 0.1°C, fluctuation ±0.3°C at 37°C
CO₂ control 0-20%, 0.1% resolution, imported infrared (IR) sensor
Heating method Direct heat (air jacket), six-face heating elements
Controller 7-inch touchscreen with real-time temp/CO₂ curves, USB CSV/PDF export
Chamber 304 stainless steel, rounded corners, coved floor
Humidification Water dish (standard); high RH maintained around open plates
Gas fitting SMC solenoid valve; standard CO₂ cylinder regulator (pure or premix)
Decontamination options HEPA filtration · UV-C lamp (programmable cycle)
Power 220V/50Hz standard; 110V/60Hz on request
Certification CE marked; built under ISO 9001; sensor calibration certificate supplied

Where you would install it

  • Mammalian cell culture. adherent and suspension lines at 37°C / 5% CO₂ with stable medium pH
  • Tissue culture. primary isolation and scaffold work with fast recovery after access
  • Cell-line maintenance. long unattended runs backed by 1-minute logging and alarms
  • Shared culture rooms. 160L capacity with per-user shelf zones and decontamination options

Frequently asked questions

What sizes does the BPN come in?

80L and 160L. Both share the same direct heat platform, IR sensor, touchscreen and data export; the 160L adds shelf capacity for multi-user rooms.

What CO₂ range does it control?

0-20% with 0.1% resolution, measured by an imported infrared sensor and metered through an SMC solenoid valve. Standard cell culture setpoint is 5%.

What is the temperature uniformity?

±0.3°C fluctuation at 37°C over the RT+5~60°C range, with 0.1°C resolution on the touchscreen.

Does it have data export?

Yes — the touchscreen logs temperature and CO₂ at one-minute intervals, displays live curves, and exports CSV or PDF over USB.

Is there a HEPA filter?

HEPA filtration of the chamber atmosphere is available, alongside the programmable UV-C decontamination lamp and autoclave-compatible shelves.

Additional information

Capacity

160L, 80L

Voltage

220V/50Hz

Chamber Material

304 Stainless

Control Type

7-inch Touchscreen

Temperature Range

RT+5~60°C

Humidity Control

Yes

Lighting Type

No

Certification

CE, ISO 9001, RoHS

Incubator Type

CO₂

Cooling System

No Cooling (Heated Only)

Heating Technology

Direct Heat

Airflow Type

Forced Air

Form Factor

Benchtop, Floor Standing

Door Type

Inner Glass Door

Decontamination

UV-C, HEPA Filter

Compliance

ISO 9001

Reviews

There are no reviews yet.

Be the first to review “WSHLAB CO₂ Incubator”

Your email address will not be published. Required fields are marked *

Technical documents

Datasheets, certificates and safety documents are available from our product specialists.

Request documentation